Název: Strain-stress analysis of lower limb with applied fixator
Autoři: Mrázek, Michal
Florian, Zdeněk
Veselý, Radek
Borák, Libor
Citace zdrojového dokumentu: Applied and Computational Mechanics. 2010, vol. 4, no. 1, p. 67-78.
Datum vydání: 2010
Nakladatel: University of West Bohemia
Typ dokumentu: článek
article
URI: http://www.kme.zcu.cz/acm/index.php/acm/article/view/54/37
http://hdl.handle.net/11025/1391
ISSN: 1802-680X (Print)
2336-1182 (Online)
Klíčová slova: deformačně-napěťová analýza;zevní fixace;holenní kost
Klíčová slova v dalším jazyce: strain-stress analysis;external fixation;tibia
Abstrakt: This paper compares physiological state of tibia before and after application of an external fixator. The fixator systems’ models but also model of tibia are loaded in the direction of body axis. The paper is focused on the examination of differences in stiffness before and after the application of fixation. Two types of axial external fixators are compared. Both fixators differ in their construction. The first fixator is two-frame and fixation rods are used for fixing the bone tissue (variant I). The second one is fixed into tibia with screws (variant II). We have found out that the two-frame external fixator has much bigger stiffness during limb fixation than the fixator with one body. Much higher deformations compared to physiological state of tibia occur in the variant II.
Práva: © 2010 University of West Bohemia. All rights reserved.
Vyskytuje se v kolekcích:Volume 4, number 1 (2010)
Volume 4, number 1 (2010)

Soubory připojené k záznamu:
Soubor Popis VelikostFormát 
37.pdf1,22 MBAdobe PDFZobrazit/otevřít


Použijte tento identifikátor k citaci nebo jako odkaz na tento záznam: http://hdl.handle.net/11025/1391

Všechny záznamy v DSpace jsou chráněny autorskými právy, všechna práva vyhrazena.