dc.contributor.authorTeodorescu, Mike
dc.contributor.authorTeodorescu, Horia-Nicolai
dc.contributor.editorPinker, Jiří
dc.date.accessioned2020-11-05T14:15:14Z
dc.date.available2020-11-05T14:15:14Z
dc.date.issued2020
dc.identifier.citation2020 International Conference on Applied Electronics: Pilsen, 8th – 9h September 2020, Czech Republic.en
dc.identifier.isbn978-80-261-0891-7 (Print)
dc.identifier.isbn978-80-261-0892-4 (Online)
dc.identifier.issn1803-7232 (Print)
dc.identifier.issn1805-9597 (Online)
dc.identifier.urihttp://hdl.handle.net/11025/39928
dc.format6 s.cs
dc.format.mimetypeapplication/pdf
dc.language.isoenen
dc.publisherZápadočeská univerzita v Plznics
dc.rights© Západočeská univerzita v Plznics
dc.subjectřízenícs
dc.subjectregulace tlaku vzduchucs
dc.subjectopen designcs
dc.subjectnízký výkoncs
dc.subjectnelineární charakteristikacs
dc.subjectpřenosné zařízenícs
dc.subjectPPEcs
dc.subjectPAPRcs
dc.subjectCOVID-19cs
dc.titleAn Open Design for an Air Flow and Pressure Control Circuit for Medical Devicesen
dc.typeconferenceObjecten
dc.typekonferenční příspěvekcs
dc.rights.accessopenAccessen
dc.type.versionpublishedVersionen
dc.description.abstract-translatedWe report on an air control system with low power consumption for small enclosures with mainly medical applications, such as lung ventilators, APAP machines, PAPR isolation hoods, surgical applications, and aerosol containment. The circuit shared can be built from cost effective, commercially available parts; it represents a pressure control system for positive pressure environments such as hoods used in personal protection equipment (PPE), as proposed in today’s COVID-19 epidemic. An essential feature of the approach is that the circuit uses advanced off the shelf modules that include microcontrollers, but the overall design remains analogic. Another contribution is to derive rigorous design equations that bring together the pneumatic and electric nonlinear equations describing an inflatable enclosure, the linearized pneumatic circuits, the power device (fan), and the control loop. We warn that the designs in this paper are for scientific testing only and not for any medical use; no guarantee is implied. See the Warnings and Disclaimer.en
dc.subject.translatedcontrolen
dc.subject.translatedair pressure controlen
dc.subject.translatedopen designen
dc.subject.translatedlow poweren
dc.subject.translatednonlinear characteristicen
dc.subject.translatedportable equipmenten
dc.subject.translatedPPEen
dc.subject.translatedPAPRen
dc.subject.translatedCOVID-19en
dc.type.statusPeer-revieweden
Appears in Collections:Applied Electronics 2020
Applied Electronics 2020

Files in This Item:
File Description SizeFormat 
09232840.pdfPlný text910,56 kBAdobe PDFView/Open


Please use this identifier to cite or link to this item: http://hdl.handle.net/11025/39928

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.