Full metadata record
DC poleHodnotaJazyk
dc.contributor.authorGeorgiev, Vjačeslav
dc.contributor.authorVavroch, Ondřej
dc.contributor.authorZich, Jan
dc.contributor.authorNožka, Libor
dc.contributor.editorPinker, Jiří
dc.date.accessioned2020-11-05T09:23:13Z
dc.date.available2020-11-05T09:23:13Z
dc.date.issued2020
dc.identifier.citation2020 International Conference on Applied Electronics: Pilsen, 8th – 9h September 2020, Czech Republic.en
dc.identifier.isbn978-80-261-0891-7 (Print)
dc.identifier.isbn978-80-261-0892-4 (Online)
dc.identifier.issn1803-7232 (Print)
dc.identifier.issn1805-9597 (Online)
dc.identifier.urihttp://hdl.handle.net/11025/39904
dc.format4 s.cs
dc.format.mimetypeapplication/pdf
dc.language.isoenen
dc.publisherZápadočeská univerzita v Plznics
dc.rights© Západočeská univerzita v Plznics
dc.subjectFPGAcs
dc.subjectfyzika vysokých energiícs
dc.subjectPMTcs
dc.subjectfotonásobičcs
dc.subjectLED ovladačcs
dc.subjectfyzické vybavenícs
dc.subjectPWMcs
dc.subjectradiační kalenícs
dc.titleDiagnostic Device for Photomultiplier Tubes at ARP ToF Detectoren
dc.typeconferenceObjecten
dc.typekonferenční příspěvekcs
dc.rights.accessopenAccessen
dc.type.versionpublishedVersionen
dc.description.abstract-translatedLong-term usage of photomultiplier tubes (PMTs) led to the discovery of aging effects especially in harsh environment as a radiation is. Aging decreases the efficiency of the PMTs. This means PMTs need to the recalibration after some time. Because CERN′s Large Hadron Collider (LHC) runs for few years, the diagnostics of the PMTs is necessary. Eventual recalibration without removing the PMT is possible to accomplish by integrating the light source nearby PMT′s chassis. The light source should by as precise as possible. The used LED source needs a driver. The driver consists of controllable power source and timing controller. The perfect timing of the pulses is controlled by FPGA. An attention to the radiation hardened parts was needed because of the device proximity to the radiation source. The development and testing of the PMT diagnostic device (PDD) are described in this paper.en
dc.subject.translatedFPGAen
dc.subject.translatedhigh energy physicsen
dc.subject.translatedPMTen
dc.subject.translatedphotomultiplieren
dc.subject.translatedLED driveren
dc.subject.translatedphysical instrumentationen
dc.subject.translatedPWMen
dc.subject.translatedrad-harden
dc.type.statusPeer-revieweden
Vyskytuje se v kolekcích:Konferenční příspěvky / Conference Papers (KAE)
Applied Electronics 2020