Title: Calibration of Instrument Current Transformer Test Sets
Authors: Draxler, Karel
Hlaváček, Jan
Styblíková, Renata
Citation: 2019 International Conference on Applied Electronics: Pilsen, 10th – 11th September 2019, Czech Republic, p.43-46.
Issue Date: 2019
Publisher: Západočeská univerzita v Plzni
Document type: konferenční příspěvek
conferenceObject
URI: http://hdl.handle.net/11025/35511
ISBN: 978–80–261–0812–2 (Online)
978–80–261–0813–9 (Print)
ISSN: 1803–7232 (Print)
1805-9597 (Online)
Keywords: přístrojový proudový transformátor;kalibrace;poměrová chyba;fázový posun
Keywords in different language: instrument current transformer;calibration;ratio error;phase displacement
Abstract in different language: The paper presents the basic layout for instrument current transformer calibration using comparative method with a standard. The error difference is evaluated using electronic transformer test sets. Calibrating procedure of these systems using error simulation is also described. In conclusion, results obtained using described procedure are presented.
Rights: © Západočeská univerzita v Plzni
Appears in Collections:Applied Electronics 2019
Applied Electronics 2019

Files in This Item:
File Description SizeFormat 
Draxler.pdfPlný text134,54 kBAdobe PDFView/Open


Please use this identifier to cite or link to this item: http://hdl.handle.net/11025/35511